Abstract
Recently, new high-resolution cadmium–zinc–telluride (CZT) drift strip detectors for room temperature gamma-ray spectroscopic imaging were developed by our group. The CZT detectors equipped with orthogonal anode/cathode collecting strips, drift strips and dedicated pulse processing allow a detection area of 6 × 20 mm2 and excellent room temperature spectroscopic performance (0.82% FWHM at 661.7 keV). In this work, we investigated the potentialities of these detectors for prompt gamma-ray spectroscopy (PGS) in boron neutron capture therapy (BNCT). The detectors, exploiting the measurement of the 478 keV prompt gamma rays emitted by 94%7Li nuclides from the10B(n, α)7Li reaction, are very appealing for the development of single-photon emission computed tomography (SPECT) systems and Compton cameras in BNCT. High-resolution gamma-ray spectra from10B samples under thermal neutrons were measured at the T.R.I.G.A. Mark II research nuclear reactor of the University of Pavia (Italy).
Lingua originale | English |
---|---|
Numero di pagine | 11 |
Rivista | Sensors |
Volume | 22 |
Stato di pubblicazione | Published - 2022 |
All Science Journal Classification (ASJC) codes
- ???subjectarea.asjc.1600.1602???
- ???subjectarea.asjc.1700.1710???
- ???subjectarea.asjc.3100.3107???
- ???subjectarea.asjc.1300.1303???
- ???subjectarea.asjc.3100.3105???
- ???subjectarea.asjc.2200.2208???