Portable CdTe detection system for mammographic X-ray spectroscopy

Leonardo Abbene, Giuseppe Raso, Gaetano Gerardi, Francesco Fauci, Stumbo, Raso, L. Barone Tonghi, Del Sordo, Gerardi, Francesco Fauci

Risultato della ricerca: Articlepeer review

7 Citazioni (Scopus)

Abstract

This paper describes a portable apparatus to be utilized in mammographic X-ray spectroscopy under clinical conditions. The system, based on a CdTe solid-state detector, is able to directly measure mammographic X-ray tube spectra. Good system response to monoenergetic photons was measured using X-ray and γ-ray calibration sources (109Cd and 241Am). The measured molybdenum X-ray spectra, in agreement with simulated spectra, show the good spectral capability of the system also at high photon fluence rates, as typical of clinical mammography. Low tailing, no secondary X-ray escape and low pile-up distortions in the measured spectra indicate that this portable system is suitable for mammographic X-ray spectroscopy. © Società Italiana di Fisica.