FEM and experimental analysis of a total knee prosthesis

Risultato della ricerca: Articlepeer review

Abstract

In this study, a comparison between two different approaches used to study a total knee prosthesis is presented. In particular, the contact area of the components of knee prosthesis has been evaluated using both a numerical and an experimental approach. The numerical analysis has been performed by FEM Models, whereas the experimental study has been conducted using an ultrasonic-based method. To setup the FEM simulations, CAD Models of the components of the prosthesis have been reconstructed using a classic reverse engineering approach. Obtained results has allowed evaluating the contact area of the components of the prosthesis and demonstrated a very good level of correlation between numerical and experimental data.
Lingua originaleEnglish
pagine (da-a)1718-1724
Numero di pagine7
RivistaJournal of Engineering and Applied Sciences
Volume13
Stato di pubblicazionePublished - 2018

All Science Journal Classification (ASJC) codes

  • ???subjectarea.asjc.2200.2200???

Fingerprint

Entra nei temi di ricerca di 'FEM and experimental analysis of a total knee prosthesis'. Insieme formano una fingerprint unica.

Cita questo