Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields

M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro

Risultato della ricerca: Paper

Abstract

Alongside with the development of Neutron Capture Therapy (NCT) and the use of thermal neutrons for radiotherapeutic purposes, many efforts have been devoted to the characterization of the beam in order to optimize therapy procedures. Reliable dose measurements should be able to determine the various (neutrons and photonic) components of the mixed beam usually employed for therapy. This paper studies the effect of additives such as boric and gadolinium nuclei on the sensitivity of neutron organic (alanine and ammonium tartrate) dosimeters analyzed through Electron Spin Resonance (ESR) technique (Marrale, 2014). These dosimeters were exposed to a mixed (neutron-gamma) field mainly composed of thermal neutrons. The choice of 10B and 64Gd as nuclei additives is due to their very high capture cross section for thermal neutrons. Also, after the nuclear reaction with thermal neutrons are emitted particles, which in turn release their energy in the vicinity of the reaction site (Marrale, 2008). The irradiation with mixed field (neutron-gamma) were performed within the thermal column of the TRIGA reactor, University of Pavia. Dosimeters readout was performed through the Electron Spin Resonance spectrometer Bruker ECS106 located at the Laboratory of Dosimetry ESR / TL of the Department of Physics and Chemistry - University of Palermo. We found that the addition of Gadolinium allows to largely increase the sensitivity of the dosimeters for thermal neutrons. In particular, a low concentration (5% by weight) of gadolinium oxide leads to an improvement of the sensitivity of neutrons more than 10 times. In addition, for this low content of gadolinium the photon tissue equivalence is not heavily reduced. This experimental analyses are compared with computational analyses carried out by means of Monte Carlo simulations performed with the MCNP (Monte Carlo N-Particle) transport code. A good agreement was observed for alanine dosimeters.
Lingua originaleEnglish
Stato di pubblicazionePublished - 2016

Fingerprint

alanine
organic compounds
dosimeters
thermal neutrons
electron paramagnetic resonance
gadolinium
neutrons
therapy
sensitivity
lead oxides
nuclei
nuclear reactions
absorption cross sections
equivalence
readout
low concentrations
reactors
photonics
spectrometers
chemistry

Cita questo

M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro (2016). Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields.

Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields. / M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro.

2016.

Risultato della ricerca: Paper

M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro 2016, 'Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields'.
M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro. Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields. 2016.
M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro. / Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields.
@conference{e0a70bee8a7b4d558ae3cba5c9dc2e04,
title = "Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields",
abstract = "Alongside with the development of Neutron Capture Therapy (NCT) and the use of thermal neutrons for radiotherapeutic purposes, many efforts have been devoted to the characterization of the beam in order to optimize therapy procedures. Reliable dose measurements should be able to determine the various (neutrons and photonic) components of the mixed beam usually employed for therapy. This paper studies the effect of additives such as boric and gadolinium nuclei on the sensitivity of neutron organic (alanine and ammonium tartrate) dosimeters analyzed through Electron Spin Resonance (ESR) technique (Marrale, 2014). These dosimeters were exposed to a mixed (neutron-gamma) field mainly composed of thermal neutrons. The choice of 10B and 64Gd as nuclei additives is due to their very high capture cross section for thermal neutrons. Also, after the nuclear reaction with thermal neutrons are emitted particles, which in turn release their energy in the vicinity of the reaction site (Marrale, 2008). The irradiation with mixed field (neutron-gamma) were performed within the thermal column of the TRIGA reactor, University of Pavia. Dosimeters readout was performed through the Electron Spin Resonance spectrometer Bruker ECS106 located at the Laboratory of Dosimetry ESR / TL of the Department of Physics and Chemistry - University of Palermo. We found that the addition of Gadolinium allows to largely increase the sensitivity of the dosimeters for thermal neutrons. In particular, a low concentration (5{\%} by weight) of gadolinium oxide leads to an improvement of the sensitivity of neutrons more than 10 times. In addition, for this low content of gadolinium the photon tissue equivalence is not heavily reduced. This experimental analyses are compared with computational analyses carried out by means of Monte Carlo simulations performed with the MCNP (Monte Carlo N-Particle) transport code. A good agreement was observed for alanine dosimeters.",
author = "{M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro} and Tomarchio, {Elio Angelo} and Anna Longo and Maurizio Marrale and Salvatore Gallo and Salvatore Panzeca",
year = "2016",
language = "English",

}

TY - CONF

T1 - Electron Spin Resonance dosimetry using organic compounds (alanine and ammonium tartrate) for mixed neutron-gamma fields

AU - M. Marrale; S. Gallo; A. Longo; S. Panzeca; D. Dondi; E. Tomarchio; A. Zeffiro

AU - Tomarchio, Elio Angelo

AU - Longo, Anna

AU - Marrale, Maurizio

AU - Gallo, Salvatore

AU - Panzeca, Salvatore

PY - 2016

Y1 - 2016

N2 - Alongside with the development of Neutron Capture Therapy (NCT) and the use of thermal neutrons for radiotherapeutic purposes, many efforts have been devoted to the characterization of the beam in order to optimize therapy procedures. Reliable dose measurements should be able to determine the various (neutrons and photonic) components of the mixed beam usually employed for therapy. This paper studies the effect of additives such as boric and gadolinium nuclei on the sensitivity of neutron organic (alanine and ammonium tartrate) dosimeters analyzed through Electron Spin Resonance (ESR) technique (Marrale, 2014). These dosimeters were exposed to a mixed (neutron-gamma) field mainly composed of thermal neutrons. The choice of 10B and 64Gd as nuclei additives is due to their very high capture cross section for thermal neutrons. Also, after the nuclear reaction with thermal neutrons are emitted particles, which in turn release their energy in the vicinity of the reaction site (Marrale, 2008). The irradiation with mixed field (neutron-gamma) were performed within the thermal column of the TRIGA reactor, University of Pavia. Dosimeters readout was performed through the Electron Spin Resonance spectrometer Bruker ECS106 located at the Laboratory of Dosimetry ESR / TL of the Department of Physics and Chemistry - University of Palermo. We found that the addition of Gadolinium allows to largely increase the sensitivity of the dosimeters for thermal neutrons. In particular, a low concentration (5% by weight) of gadolinium oxide leads to an improvement of the sensitivity of neutrons more than 10 times. In addition, for this low content of gadolinium the photon tissue equivalence is not heavily reduced. This experimental analyses are compared with computational analyses carried out by means of Monte Carlo simulations performed with the MCNP (Monte Carlo N-Particle) transport code. A good agreement was observed for alanine dosimeters.

AB - Alongside with the development of Neutron Capture Therapy (NCT) and the use of thermal neutrons for radiotherapeutic purposes, many efforts have been devoted to the characterization of the beam in order to optimize therapy procedures. Reliable dose measurements should be able to determine the various (neutrons and photonic) components of the mixed beam usually employed for therapy. This paper studies the effect of additives such as boric and gadolinium nuclei on the sensitivity of neutron organic (alanine and ammonium tartrate) dosimeters analyzed through Electron Spin Resonance (ESR) technique (Marrale, 2014). These dosimeters were exposed to a mixed (neutron-gamma) field mainly composed of thermal neutrons. The choice of 10B and 64Gd as nuclei additives is due to their very high capture cross section for thermal neutrons. Also, after the nuclear reaction with thermal neutrons are emitted particles, which in turn release their energy in the vicinity of the reaction site (Marrale, 2008). The irradiation with mixed field (neutron-gamma) were performed within the thermal column of the TRIGA reactor, University of Pavia. Dosimeters readout was performed through the Electron Spin Resonance spectrometer Bruker ECS106 located at the Laboratory of Dosimetry ESR / TL of the Department of Physics and Chemistry - University of Palermo. We found that the addition of Gadolinium allows to largely increase the sensitivity of the dosimeters for thermal neutrons. In particular, a low concentration (5% by weight) of gadolinium oxide leads to an improvement of the sensitivity of neutrons more than 10 times. In addition, for this low content of gadolinium the photon tissue equivalence is not heavily reduced. This experimental analyses are compared with computational analyses carried out by means of Monte Carlo simulations performed with the MCNP (Monte Carlo N-Particle) transport code. A good agreement was observed for alanine dosimeters.

UR - http://hdl.handle.net/10447/204622

UR - http://eventi.cnism.it/materials2016/submission/view/1414

M3 - Paper

ER -