Carotid intima-media thickness measurement through semi-automated detection software and analysis of vascular walls

    Risultato della ricerca: Article

    6 Citazioni (Scopus)

    Abstract

    AIM: The aim of the present study was to compare the semi-automatic measurement of carotid intima thickness (RFQIMT - Esaote, Italy), with the conventional method.METHODS: We enrolle 81 patients, mean age 46 years ±15, with no history of cardio-cerebrovascular events and we assessed the traditional cardiovascular risk factors. We examined the IMT of the common carotid artery with manual and RFQIMT method (based on the "Radio Frequency" signal), according to the ASE protocol.RESULTS: Semi-automatic measurement was on average lower than manual measurement (617 μm ±191 vs. 676 μm ± 222) with a statistically significant difference (P<0.01). In agreement with manual measurements, the values of RFQIMT increased with increasing age and presence of cardiovascular risk factors.CONCLUSION: The RFQIMT measurement was, on average, lower than manual measurement, this means that probably, age- and gender-related reference values of RFQIMT need to be revised.
    Lingua originaleEnglish
    pagine (da-a)349-353
    Numero di pagine5
    RivistaINTERNATIONAL ANGIOLOGY
    Volume32
    Stato di pubblicazionePublished - 2013

      Fingerprint

    All Science Journal Classification (ASJC) codes

    • Cardiology and Cardiovascular Medicine

    Cita questo